1. OBJECTIVE
- 1 To lay down a procedure for adjustment of chromatographic conditions for System suitability
2. RESPONSIBILITY
2.1 Chemist / Executive –QC – To follow the SOP.
2.2 In Charge / His Designee – QC – To ensure the compliance of SOP
3. PROCEDURE
3.1 The various parameters of a chromatographic conditions shall be adjusted to satisfy the system suitability criteria for particular analysis
3.2 The parameters shall be adjusted in order to ensure the separation required for satisfactory performance of the test or assay without fundamentally modifying the method
3.3 The adjustment of critical parameters shall be performed if they are clearly mentioned in the monograph in order to achieve the system suitability
3.4 The adjusting limits of parameters shall be followed as mentioned in annexure-1
3.5 Adjustments shall be permitted only if suitable standards (including reference standards) are available for all compounds used in the suitability test and are used to show that the adjustments have improved the quality of the chromatography in meeting system suitability requirements
3.6 Adjustments of Multiple parameters shall be avoided in order to avoid cumulative effect on the performance of the method.
3.7 The modified parameters shall be recorded in worksheet and method shall be validated.
- Adjustments to chromatographic systems performed in order to comply with System suitability requirements should not be made to compensate for column failure or to circumvent replacing a deteriorated column.
- Adjustment of the various parameters in reverse phase liquid chromatographic methods will not always result in satisfactory chromatography. In that case, it may be necessary to change the column with another of the same type but from a different manufacturer.
4. ABBREVIATIONS
4.1 QC – Quality Control.
5. REFERENCES
- British Pharmacopoeia
- United states pharmacopoeia forum
6. ANNEXURES
6.1 Annexure-1 Limits for adjusting parameters for system suitability
ANNEXURE – 1
Limits for Adjusting Parameters for System Suitability
Parameters | Thin Layer chromatography | High performance Liquid chromatography | Gas chromatography |
Composition of Mobile Phase. | The amount of minor component may be adjusted by ± 30% relative or ±2% absolute, whichever is larger .No other component is altered by more than 10 % absolute | The amount of minor component may be adjusted by ± 30% relative or ±2% absolute, whichever is larger .No other component is altered by more than 10 % absolute | Not Applicable |
pH of mobile phase | ±0.2 pH, unless otherwise stated in the monograph or ± 0.1 pH for neutral substances | ±0.2 pH, unless otherwise stated in the monograph or ± 0.1 pH for neutral substances | Not Applicable |
Concentration of salts in the buffer component of Mobile phase | ± 10% | ± 10% | Not Applicable |
Detector Wavelength | Not Applicable | No adjustment permitted | Not Applicable |
Stationary phase | As Mentioned in monograph | Column length : ± 70 % Column I.D : ± 50 % Particle size : Maximal reduction of 50 % No increase permitted | Column length : ± 70 % Column I.D : ± 50 % Particle size : Maximal reduction of 50 %,No increase Permitted Film Thickness – 50 % to 100% |
Flow Rate | Not Applicable | ± 50 % | ± 50 % |
Parameters | Thin Layer chromatography | High performance Liquid chromatography | Gas chromatography |
Temperature | Not Applicable | ± 10 % , to a maximum of 60° C | ± 10 % |
Injection volume / Application volume | Decreased to 20 % of the prescribed Volume if using fine particle size plates ( 2-10 micron) | May be decreased, provided detection and repeatability of the peaks to be determined satisfactory | May be decreased, provided detection and repeatability of the peaks to be determined satisfactory |
Column temperature | Not Applicable | Can be adjusted by as much as ± 20°. Column thermo stating to improve control and reproducibility of retention time | Can be adjusted by as much as ± 2 % in terms of absolute temperature |
Oven temperature program | Not Applicable | Not Applicable | Can be adjusted by as much as ± 2 % in terms of absolute temperature. For the times specified for the temperature to be maintained or for the temperature to be changed from one to another, an adjusted of upto ± 20% is permitted |